DocumentCode
432166
Title
Cavitation enhances treatment depth when combined with thermal effect using a plane ultrasound transducer: an in vivo study
Author
Goldenstedt, C. ; Melodelima, D. ; Mithieux, F. ; Chesnais, S. ; Theillére, Y. ; Cathignol, D.
Author_Institution
INSERM Unit 556, Lyon, France
Volume
1
fYear
2004
fDate
23-27 Aug. 2004
Firstpage
709
Abstract
Minimally invasive techniques using miniaturised ultrasound applicators have been developed for surgical and medical applications. However, the volume of treatment remains limited. Therefore, we propose to combine cavitation with a thermal effect to enhance the treatment depth. This study was conducted in vivo on pig liver, 51 exposures were performed on the surface of the liver, with a miniaturised applicator using a plane transducer (3×10 mm2) operating at 4.75 MHz. This study aimed to compare two methods of treatment. The first one produced the thermal effect exclusively. The second method used cavitation in addition to the thermal effect. Cavitation bubbles were initiated along the beam axis by a 60 W/cm2 pulse (0.5 s) followed by a continuous exposure, which generated the thermal effect. The method using cavitation produced the deepest lesions. At an intensity of 30 W/cm2 applied for 30 s, the lesions induced with cavitation bubbles were 1.7 times deeper than those produced without cavitation bubbles. Histological analysis shows that they correspond to coagulation necroses. Thus, these results show that the initiation of cavitation bubbles prior to the exposure increases the therapeutic depth significantly. Consequently, this method might open the way to the treatment of large diameter tumours.
Keywords
biomedical ultrasonics; bubbles; cavitation; coagulation; hyperthermia; radiation therapy; surgery; tumours; ultrasonic transducers; 0.5 s; 10 mm; 3 mm; 30 s; 4.75 MHz; cavitation bubbles; coagulation necroses; histological analysis; large diameter tumours; lesions; medical applications; miniaturised ultrasound applicators; plane ultrasound transducer; surgical applications; thermal effect; treatment depth; Applicators; Biomedical equipment; Biomedical transducers; In vivo; Lesions; Liver; Medical services; Minimally invasive surgery; Ultrasonic imaging; Ultrasonic transducers;
fLanguage
English
Publisher
ieee
Conference_Titel
Ultrasonics Symposium, 2004 IEEE
ISSN
1051-0117
Print_ISBN
0-7803-8412-1
Type
conf
DOI
10.1109/ULTSYM.2004.1417820
Filename
1417820
Link To Document