• DocumentCode
    432224
  • Title

    Multilayer PMN-PT single crystal transducer for medical application

  • Author

    Rhim, Sung Min ; Jung, Ho ; Lee, Ki-Jong

  • Author_Institution
    Humanscan Co. Ltd., Sunggok-Dong, South Korea
  • Volume
    2
  • fYear
    2004
  • fDate
    23-27 Aug. 2004
  • Firstpage
    1021
  • Abstract
    The performance of multilayer 64 channel 3.5 MHz phased array ultrasonic probe using 0.67Pb(Mg13/Nb23/)O3-0.33PbTiO3 single crystal was investigated. Multilayer PMN-PT single crystal with 2 layers shows 5 times larger dielectric constant and slightly higher k33´ than single layer. 2 way sensitivity of the multilayer phased array is about 4 dB higher than that of single layer phased array. The -6 dB bandwidth of multilayer phased array is about 100% and this is similar to that of single layer phased array. Sensitivity variation between single and multilayer phased array is similar and the deviation is almost same as the PZT ceramic phased array.
  • Keywords
    biomedical ultrasonics; lead compounds; multilayers; permittivity; probes; ultrasonic transducer arrays; 3.5 MHz; medical application; multilayer PMN-PT; performance; phased array ultrasonic probe; single crystal transducer; Biomedical equipment; Biomedical transducers; Dielectric constant; Medical services; Niobium; Nonhomogeneous media; Phased arrays; Probes; Ultrasonic transducer arrays; Ultrasonic transducers;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics Symposium, 2004 IEEE
  • ISSN
    1051-0117
  • Print_ISBN
    0-7803-8412-1
  • Type

    conf

  • DOI
    10.1109/ULTSYM.2004.1417943
  • Filename
    1417943