• DocumentCode
    432355
  • Title

    Evaluation of 2-D speckle tracking based strain rate imaging (SRI) using a 3-D heart simulation model

  • Author

    Chen, X. ; Li, X. ; Sahn, D. ; Kim, K. ; Xie, H. ; O´Donnell, M.

  • Author_Institution
    Biomed. Eng. Dept., Michigan Univ., Ann Arbor, MI, USA
  • Volume
    3
  • fYear
    2004
  • fDate
    23-27 Aug. 2004
  • Firstpage
    2125
  • Abstract
    Preliminary 2D SRI results from a porcine heart model showed significant average residual accumulation error (RAE) in the total displacement, similar to widely reported baseline drift artifacts in SRI. Therefore, to understand the performance of 2D SRI methods in a controlled 3D environment, a left-ventricle simulation model with physiological deformation was developed. Torsion was added to move tissue in and out of the image plane. A phased-array imaging system was simulated to generate 2D image sequences for typical cardiac scans. Results show the presence of RAE, which reaches a minimum at a frame rate (FR) of 75 frame/cycle corresponding to 2-3% peak frame-to-frame strain. RAE increases as torsion increases. In addition. the RAE map also spatially correlates well with both out-of-plane motion and cross-correlation coefficient maps. These results suggest that out-of-plane motion plays a significant role in RAE. To optimize SNR in displacement estimates a retrospective processing strategy employing different frame intervals for different periods in a cardiac cycle is presented.
  • Keywords
    acoustic correlation; biological tissues; echocardiography; image sequences; medical image processing; speckle; torsion; tracking; ultrasonic imaging; 2D image sequences; 2D speckle tracking; 3D heart simulation model; RAE; SNR; SRI; average residual accumulation error; cardiac cycle; cross-correlation coefficient maps; frame intervals; left-ventricle simulation model; out-of-plane motion; peak frame-to-frame strain; phased-array imaging system; physiological deformation; porcine heart model; retrospective processing; spatial correlation; strain rate imaging; tissue; torsion; Additive noise; Biomedical imaging; Capacitive sensors; Decorrelation; Deformable models; Displays; Heart; Signal to noise ratio; Speckle; Tin;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Ultrasonics Symposium, 2004 IEEE
  • ISSN
    1051-0117
  • Print_ISBN
    0-7803-8412-1
  • Type

    conf

  • DOI
    10.1109/ULTSYM.2004.1418257
  • Filename
    1418257