Abstract :
Gradient coils for magnetic resonance imaging (MRI) require large currents (> 500 A) for the gradient field strength, as well as high voltage (> 1600 V) for fast slew rates. Additionally, extremely high fidelity, reproducing the command signal, is critical for image quality. A new driver topology recently proposed can provide the high power and operate at high switching frequency allowing high bandwidth control. The paper presents additional improvements to the new driver architecture, and more importantly, describes the digital control design and implementation, crucial to achieve the required performance level. The power stage and control have been build and tested with the experimental results showing that the performance achieved with the new digital control capability, more than fulfills the system requirements
Keywords :
coils; digital control; insulated gate bipolar transistors; magnetic resonance imaging; IGBT; MRI; bandwidth control; digital control; gradient coil; gradient driver; image quality; magnetic resonance imaging; slew rate; switching frequency; Bandwidth; Coils; Control systems; Digital control; Image quality; Magnetic resonance imaging; Switching frequency; System testing; Topology; Voltage;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Applied Power Electronics Conference and Exposition, 2005. APEC 2005. Twentieth Annual IEEE
Conference_Location :
Austin, TX
Print_ISBN :
0-7803-8975-1
Type :
conf
DOI :
10.1109/APEC.2005.1453131
Filename :
1453131
Link To Document :
بازگشت