Abstract :
The ICC neutron detector is an ionisation chamber with three neutron absorbing converter electrodes of 6Li metal foils. On both sides the foils face collector electrodes that are positively biased. The neutron detection was investigated at the pulsed neutron source of the Argonne National Laboratory. The induced charge due to drifting electrons that are generated by the energetic 3H recoil nucleus in the C2F6 gas between the electrodes was measured as function of gas pressure, gas contamination, and applied drift voltage. The neutron detection efficiency was determined and the gamma insensitivity was estimated. The pulse heights were calibrated by the induced charge from test pulses. The timing characteristics of the detector pulses were assessed, and TOF measurements demonstrated the usefulness of the detector. First neutron diffraction experiments were made with structured collection electrodes for position-sensitive detection and for estimating the achievable spatial resolution.
Keywords :
drift chambers; electrodes; helium-3 counters; ionisation chambers; neutron detection; neutron sources; position sensitive particle detectors; 6Li converters; 6Li metal foils; C2F6 gas; ICC neutron detector; TOF measurements; detector pulses; drift voltage; drifting electron induced charge; energetic 3H recoil nucleus; gamma insensitivity; gas contamination; gas pressure; ionisation chamber; neutron absorbing converter electrodes; neutron detection efficiency; neutron diffraction experiments; position-sensitive detection; positively biased collector electrodes; pulse height calibration; pulsed neutron source; spatial resolution; test pulse induced charge; timing characteristics; Current measurement; Detectors; Electrodes; Electrons; Face detection; Ionization chambers; Laboratories; Neutrons; Pollution measurement; Pulse measurements;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium Conference Record, 2004 IEEE
ISSN :
1082-3654
Print_ISBN :
0-7803-8700-7
Electronic_ISBN :
1082-3654
Type :
conf
DOI :
10.1109/NSSMIC.2004.1462239
Filename :
1462239
Link To Document :
بازگشت