• DocumentCode
    437974
  • Title

    Design considerations for passive gamma-ray spectrometers

  • Author

    Smith, L. Eric ; Swickard, Andrea R. ; Heredia-Langner, Alejandro ; Warren, Glen A. ; Siciliano, Edward R. ; Miller, Steven D.

  • Author_Institution
    Pacific Northwest Nat. Lab., Richland, WA, USA
  • Volume
    3
  • fYear
    2004
  • fDate
    16-22 Oct. 2004
  • Firstpage
    1571
  • Abstract
    Passive gamma-ray spectrometers composed of attenuation filters and integrating detection materials provide important advantages for measurements in high-radiation environments and for long-term monitoring. Each of these applications has requirements that constrain the design of the instrument, such as incident energy range of interest, sensor size and weight, readout method, and cost. The multitude of parameters in passive spectrometer design (e.g. attenuation filter material and thickness, integrating sensor type, numbers of pixels, reconstructed energy bin structure) results in a large design space to examine. The development of generalized design optimization tools to interrogate this space and to identify promising spectrometer designs is discussed, particularly the methods used to rapidly calculate system transfer functions and the use of genetic algorithms for design optimization. Specific examples of preliminary design optimization efforts are provided.
  • Keywords
    filters; gamma-ray detection; gamma-ray spectrometers; nuclear electronics; position sensitive particle detectors; readout electronics; attenuation filter material; attenuation filter thickness; generalized design optimization tools; genetic algorithms; high-radiation environments; incident energy; integrating detection materials; integrating sensor type; long-term monitoring; passive gamma-ray spectrometers; passive spectrometer design; pixel numbers; readout method; reconstructed energy bin structure; sensor size; sensor weight; system transfer functions; Algorithm design and analysis; Attenuation measurement; Costs; Design optimization; Gamma ray detection; Gamma ray detectors; Instruments; Monitoring; Passive filters; Spectroscopy;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2004 IEEE
  • Conference_Location
    Rome
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-8700-7
  • Electronic_ISBN
    1082-3654
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2004.1462540
  • Filename
    1462540