Adjusting the low energy threshold for large bodies in PET
Author :
Turkington, T.G. ; Wilson, J.W. ; Colsher, J.G. ; Colsher, J.G.
Author_Institution :
Dept. of Radiol., Duke Univ., Durham, NC
Volume :
5
fYear :
2004
fDate :
16-22 Oct. 2004
Firstpage :
2872
Lastpage :
2876
Abstract :
The performance of a PET scanner on three different phantom sizes was studied as a function of low energy threshold (LET). Phantom cross sections ranged from 20 cm diameter circular to 28 cmtimes43 cm oval and LET´s ranged from 350 keV to 475 keV, in 25 keV increments. System sensitivity, scatter fraction, and NEC were measured over a wide range of radioactivity levels. Increasing the low energy threshold lowered both sensitivity and scatter fraction. The statistical quality of the raw data was maximized for the 425 keV setting for all three phantoms. System stability and uniformity of response was also studied for 375 keV to 450 keV thresholds, and indicated acceptable performance for this system through 425 keV
Keywords :
image scanners; phantoms; positron emission tomography; 20 to 43 cm; 350 to 475 keV; PET scanner; low energy threshold; phantom cross sections; phantom sizes; scatter fraction; system sensitivity; Detectors; Electromagnetic scattering; Energy resolution; Imaging phantoms; Medical services; National electric code; Particle scattering; Positron emission tomography; Radiology; Telephony;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium Conference Record, 2004 IEEE
Conference_Location :
Rome
ISSN :
1082-3654
Print_ISBN :
0-7803-8700-7
Electronic_ISBN :
1082-3654
Type :
conf
DOI :
10.1109/NSSMIC.2004.1466286
Filename :
1466286
Link To Document :
بازگشت