• DocumentCode
    438518
  • Title

    PET detector quality assurance using /sup 137/Cs singles transmission imaging

  • Author

    Wassenaar, R.W. ; deKemp, R.A.

  • Author_Institution
    Carleton Univ., Ottawa, Ont.
  • Volume
    5
  • fYear
    2004
  • fDate
    16-22 Oct. 2004
  • Firstpage
    2906
  • Lastpage
    2909
  • Abstract
    Daily quality assurance of the detectors, to ensure accurate efficiency calibration of the system, typically requires the acquisition of coincidence data from a flood source. In order to obtain sufficient counts, long scan times, on the order of an hour, are required. The long acquisition time and necessity of an external activity source makes such a protocol impractical for daily use. The new method has been demonstrated for the KCAT ART scanner, using the two 137Cs transmission sources. The new protocol produces a map of the counts per detector, as well as plots of the mean detector counts per block and CoV of each block. Graphical outputs allow for quick recognition of problematic areas. The factory protocol requires 90 min to obtain a 1% error on the individual detector counts. Due to the high activity of the 137Cs sources, a 1% error is obtained with an 8 min scan with the new protocol. The new protocol provides information regarding the status of all blocks, reduces acquisition time, and alleviates the need for an external activity source. The protocol presented in this study can be implemented on other 137Cs based scanners, and with coincidence flood sources if detector-specific singles histogramming is available
  • Keywords
    caesium; image scanners; positron emission tomography; radiation detection; 137Cs based scanners; 137Cs singles transmission imaging; Cs; KCAT ART scanner; PET detector quality assurance; detector-specific singles histogramming; efficiency calibration; external activity source; factory protocol; flood source; graphical outputs; individual detector counts; long acquisition time; long scan times; Calibration; Detectors; Floods; Positron emission tomography; Production facilities; Protocols; Quality assurance; Root mean square; Subspace constraints; Telephony;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2004 IEEE
  • Conference_Location
    Rome
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-8700-7
  • Electronic_ISBN
    1082-3654
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2004.1466294
  • Filename
    1466294