    Accuracy and variability of quantitative values obtained for mouse imaging using the microPET FOCUS

  • Author

    Jan, Sen ; Boisgard, R. ; Fontyn, Y. ; Eroukhmanoff, C. ; Comtat, Claude ; Trebossen, R. ; Joliot, F.

  • Author_Institution
    DRM, CEA, Orsay
  • Volume
    5
  • fYear
    2004
  • fDate
    16-22 Oct. 2004
  • Firstpage
    2934
  • Lastpage
    2937
  • Abstract
    This work is based on the optimization of quantitative analysis dedicated to the PET mouse imaging using the microPETreg FOCUS system. Five acquisition protocols were defined to improve the FOCUS sensitivity. We demonstrated that the data collection between 150-750 keV with 6 ns for coincidences windows improves the sensitivity by a factor 2 in comparison with the standard acquisition protocol (350-750 keV; 6 ns). First we showed that the inter animal variability for PET imaging using FDG, varies between 15% and 35% for liver, heart, kidney and leg muscles. These resulting values are lower than those obtained with an ex-vivo activity counting. Secondly, we demonstrated that intra animal variability is about 10% for the same organs and it is lower than the inter variability
  • Keywords
    cardiology; kidney; liver; positron emission tomography; 150 to 750 keV; 6 ns; FDG; PET mouse imaging; ex vivo activity counting; heart; interanimal variability; intraanimal variability; kidney; leg muscles; liver; microPET/FOCUS system; optimization; quantitative analysis; standard acquisition protocol; Animals; Focusing; Heart; Image analysis; Leg; Liver; Mice; Muscles; Positron emission tomography; Protocols;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2004 IEEE
  • Conference_Location
    Rome
  • ISSN
    1082-3654
  • Print_ISBN
    0-7803-8700-7
  • Electronic_ISBN
    1082-3654
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2004.1466301
  • Filename
    1466301