DocumentCode :
438528
Title :
Development of an FPGA-based data acquisition module for small animal PET
Author :
Hegyesi, G. ; Imrek, J. ; Kalinka, G. ; Molnar, J. ; Novak, D. ; Vegh, J. ; Balkay, L. ; Emri, M. ; Molnar, G. ; Tron, L. ; Bagamery, I. ; Bukki, Tamas ; Rozsa, S. ; Szabo, Zsolt ; Kerek, A.
Author_Institution :
Inst. of Nucl. Res., Hungarian Acad. of Sci., Debrecen
Volume :
5
fYear :
2004
fDate :
16-22 Oct. 2004
Firstpage :
2957
Lastpage :
2961
Abstract :
We report on the design of a DAQ module for a small animal PET camera developed at our institutes. During the design an important guideline was to develop a system which is built up from strictly identical DAQ modules, and which has no built-in hardware limitation on the maximum number of modules. The developed DAQ module comprises of an LSO scintillator crystal block, a position sensitive PMT, analog signal conditioning circuits, a digitizer, an FPGA for digital signal processing and a communication module through which the collected data is sent to a cluster of computers for post processing and storage. Instead of implementing hardware coincidence detection between the modules we attach a precise time-stamp to each event in our design, and the coincidence is determined by the data collecting computers during the post processing. The digital CFD algorithm implemented in the FPGA gives a time resolution of 2 to 3 ns FWHM for real detector signals
Keywords :
analogue circuits; data acquisition; field programmable gate arrays; gamma-ray detection; medical signal processing; nuclear electronics; position sensitive particle detectors; positron emission tomography; solid scintillation detectors; DAQ module; FPGA; LSO scintillator crystal block; analog signal conditioning circuits; communication module; data storage; digital CFD algorithm; digital signal processing; digitizer; position sensitive PMT; post processing; real detector signals; small animal PET camera; Animals; Cameras; Circuits; Data acquisition; Digital signal processing; Field programmable gate arrays; Guidelines; Hardware; Positron emission tomography; Signal processing;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium Conference Record, 2004 IEEE
Conference_Location :
Rome
ISSN :
1082-3654
Print_ISBN :
0-7803-8700-7
Electronic_ISBN :
1082-3654
Type :
conf
DOI :
10.1109/NSSMIC.2004.1466306
Filename :
1466306
Link To Document :
بازگشت