DocumentCode :
438663
Title :
New results from a test of silicon photomultiplier as readout for PET
Author :
Otte, A. Nepomuk ; Barral, Joëlle ; Dolgoshein, Boris ; Hose, Jürgen ; Klemin, Sergei ; Lorenz, Eckart ; Mirzoyan, Razmick ; Popova, Elena ; Teshima, Masahiro
Author_Institution :
Max-Planck-Inst. fur Phys., Munich, Germany
Volume :
6
fYear :
2004
fDate :
16-22 Oct. 2004
Firstpage :
3738
Abstract :
The silicon photomultiplier (SiPM) is a novel photon detector based on Geiger mode operating avalanche photodiodes. In this paper we present experimental results which show that SiPMs can be used in positron emission tomography (PET). As scintillator we use the newly developed LYSO material, which has very similar characteristics to the widely used LSO. With our setup we measure an energy resolution of (21.5±0.5)% and a coincidence time resolution of (1.51±0.07) ns, both full width at half maximum. A significant improvement in time resolution could be achieved by triggering on the first photoelectron in the signal. We also present the coincidence rate of two detector channels vs. the position of a small pointlike 22Na positronium source.
Keywords :
avalanche photodiodes; photomultipliers; positron emission tomography; positronium; silicon radiation detectors; 22Na; 22Na positronium source; APD; Geiger mode operating avalanche photodiode; PET; detector channel; lutetium yttrium orthosilicate; photoelectron triggering; photon detector; positron emission tomography; silicon photomultiplier; time resolution; Avalanche photodiodes; Detectors; Energy measurement; Energy resolution; Photomultipliers; Positron emission tomography; Signal resolution; Silicon; Testing; Time measurement;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium Conference Record, 2004 IEEE
ISSN :
1082-3654
Print_ISBN :
0-7803-8700-7
Electronic_ISBN :
1082-3654
Type :
conf
DOI :
10.1109/NSSMIC.2004.1466693
Filename :
1466693
Link To Document :
بازگشت