DocumentCode
438715
Title
Monte Carlo analysis of a mercuric iodide neutron/gamma detector
Author
Bell, Zane W. ; West, W. Geoffrey ; Pohl, Kenneth R. ; Van den Berg, Lodewijk
Author_Institution
Div. of Nucl. Sci. & Technol., Oak Ridge Nat. Lab., TN, USA
Volume
7
fYear
2004
fDate
16-22 Oct. 2004
Firstpage
4580
Abstract
The response to neutrons of a 27×27×2.6 mm Hgl2 detector covered on one face with a layer of boron was modeled with MCNP5 and MCNP-PoliMi. It was found that the detection of the 478 keV gamma ray from the 10B(n,α) and the 368 keV gamma ray from the 199Hg(n,α) could be used to estimate the average energy of incident neutrons and that these gamma rays could be used to distinguish between unmoderated radioactive and fission sources.
Keywords
Monte Carlo methods; gamma-ray detection; neutron detection; neutron-nucleus reactions; semiconductor counters; 2.6 mm; 27 mm; 10B(n,alpha); 199Hg(n,alpha); Hgl2 detector; MCNP-PoliMi; MCNP5; Monte Carlo analysis; boron layer; fission source; incident neutron energy; mercuric iodide neutron/gamma detector; unmoderated radioactive source; Aluminum; Boron; Face detection; Gamma ray detection; Gamma ray detectors; Monte Carlo methods; Neutrons; Polyethylene; Printed circuits; Telephony;