DocumentCode :
451454
Title :
Spectroscopic radiation portal monitor prototype
Author :
McCormick, Kathleen ; Stromswold, David ; Ely, James ; Schweppe, John ; Kouzes, Richard
Author_Institution :
Pacific Northwest Nat. Lab., Richland, WA
Volume :
1
fYear :
2005
fDate :
23-29 Oct. 2005
Firstpage :
292
Lastpage :
296
Abstract :
A spectroscopic radiation portal monitor (SPM) prototype consisting of four 10.16-cmtimes10.16-cmtimes40.64-cm sodium iodide (NaI) crystals has been constructed at Pacific Northwest National Laboratory (PNNL). The prototype was put through a variety of tests, including measurements of the absolute detection efficiency of unshielded sources and the detection efficiency and isotopic identification capability of the detector for shielded isotopic sources. The monitor´s response to various types of cargo and source configurations was also studied. The results of these tests are presented in this report
Keywords :
gamma-ray applications; gamma-ray detection; security; solid scintillation detectors; 10.16 cm; 40.64 cm; detection efficiency; isotopic identification; shielded isotopic sources; sodium iodide crystals; spectroscopic radiation portal monitor; Crystals; Detectors; Face detection; Portals; Prototypes; Radiation monitoring; Radioactive materials; Scanning probe microscopy; Spectroscopy; Testing;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium Conference Record, 2005 IEEE
Conference_Location :
Fajardo
ISSN :
1095-7863
Print_ISBN :
0-7803-9221-3
Type :
conf
DOI :
10.1109/NSSMIC.2005.1596256
Filename :
1596256
Link To Document :
بازگشت