DocumentCode :
451499
Title :
Virtually increasing ADC sampling frequency for high resolution timing measurements in heavy ion reactions
Author :
Bassini, R. ; Boiano, C. ; Guazzoni, P. ; Riccio, F. ; Russo, S. ; Sassi, M. ; Zetta, L.
Author_Institution :
Ist. Nazionale di Fisica Nucl., Milan
Volume :
1
fYear :
2005
fDate :
23-29 Oct. 2005
Firstpage :
565
Lastpage :
569
Abstract :
In view of performing time measurements via digital pulse shape acquisition, we have designed, built and tested a simple VME-board, able to virtually increase up to 1.6 GHz the sampling frequency of a 100 MHz analog to digital converter, without affecting its resolution. Particular care has been devoted to the codes necessary for managing the data and for the pulse height and time calibration of the board-SADC ensemble
Keywords :
analogue-digital conversion; calibration; data acquisition; ADC sampling frequency; VME-board; analog-to-digital converter; board-SADC ensemble; digital pulse shape acquisition; heavy ion reactions; high resolution timing measurements; pulse height; time calibration; Frequency conversion; Frequency measurement; Performance evaluation; Pulse measurements; Pulse shaping methods; Sampling methods; Shape measurement; Testing; Time measurement; Timing;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium Conference Record, 2005 IEEE
Conference_Location :
Fajardo
ISSN :
1095-7863
Print_ISBN :
0-7803-9221-3
Type :
conf
DOI :
10.1109/NSSMIC.2005.1596315
Filename :
1596315