Scatter estimation and motion correction in PET
Author :
Thielemans, Kris
Author_Institution :
Hammersmith Imanet Limited, UK
Volume :
3
fYear :
2005
fDate :
23-29 Oct. 2005
Abstract :
This paper discusses how to estimate scatter when performing motion correction of PET data. It concentrates on head movement, but some results are valid for non-rigid motion as well. We show that rigid motion affects scattered events differently then unscattered events. This has consequences for the scatter estimation procedure. We show with simulations, phantom measurements and patient data that our proposed method obtains fully quantitative motion-corrected PET images.
Keywords :
biomechanics; image motion analysis; medical image processing; phantoms; positron emission tomography; PET; head movement; motion correction; nonrigid motion; phantom; scatter estimation; Detectors; Electromagnetic scattering; Event detection; Head; Image reconstruction; Light scattering; Motion estimation; Particle scattering; Positron emission tomography; Sensor arrays;