DocumentCode :
451707
Title :
Continuous model of multi-pinhole SPECT devices
Author :
Rentmeester, M.C.M. ; van der Have, F. ; Beekman, F.J.
Author_Institution :
Dept. of Pharmacology & Anatomy, Univ. Medical Center Utrecht, Netherlands
Volume :
3
fYear :
2005
fDate :
23-29 Oct. 2005
Abstract :
Recently multi-pinhole SPECT systems have been proposed that allow for sub-mm resolution imaging in mice. The optimization of multi-pinhole systems using ray-tracing or Monte Carlo simulation is extremely time consuming. As an alternative we develop an approximate continuous model of a SPECT system which we apply to focused pinhole imaging of the mouse. In order to facilitate the use of analytical mathematical calculations the model assumes that the system consists of spherical layers. It uses analytical expressions of effective pinhole diameters, sensitivity, and resolution. With a pinhole-diameter adapting feedback loop that allows to compare system resolution at equal system sensitivity and vice versa the model predicts that (i) for optimal resolution pinholes should be as close as possible to the animal, (ii) with high resolution (<0.1 mm) gamma cameras a resolution improvement of at least 20% can be reached compared to a system with with standard low resolution (3.2 mm) NaI detectors in a compact system.
Keywords :
image resolution; single photon emission computed tomography; NaI detectors; focused pinhole imaging; gamma cameras; mouse; multipinhole SPECT devices; pinhole-diameter adapting feedback loop; sensitivity; sub-mm resolution imaging; Animals; Cameras; Feedback loop; Focusing; Gamma ray detection; Image resolution; Mathematical model; Mice; Predictive models; Ray tracing;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium Conference Record, 2005 IEEE
ISSN :
1095-7863
Print_ISBN :
0-7803-9221-3
Type :
conf
DOI :
10.1109/NSSMIC.2005.1596665