• DocumentCode
    456433
  • Title

    Realisation of a New MULTITASK Phantom (M.T.P) for the Verification of Global Performance of an MRI System

  • Author

    Besbes, Hatem ; Hmani, Wacel ; Essafi, Fatma ; Ben Salah, Ridha

  • Author_Institution
    Faculte de Medecine de Tunis
  • Volume
    1
  • fYear
    0
  • fDate
    0-0 0
  • Firstpage
    1153
  • Lastpage
    1158
  • Abstract
    The diversity of machines and the complexity of the technique of MRI imaging require protocols of quality control. These protocols are generally realised with a lot of phantoms. However, these operations of quality control require a lot of time. In order to reduce these operations and save time, we realise a multitask phantom. The phantom is constructed with polyamide and filled with a solution of copper sulphate (CUSO4). It is called multi-task phantom because it contains some compartment characterised by specific geometrical forms. Thus, it allows us to do several controls through only one acquisition protocol
  • Keywords
    biomedical MRI; biomedical equipment; copper compounds; quality control; MRI imaging; multitask phantom; polyamide; quality control; Biomedical imaging; Coils; Imaging phantoms; Linearity; Magnetic resonance imaging; Protocols; Protons; Quality control; Resonant frequency; Signal to noise ratio;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Information and Communication Technologies, 2006. ICTTA '06. 2nd
  • Conference_Location
    Damascus
  • Print_ISBN
    0-7803-9521-2
  • Type

    conf

  • DOI
    10.1109/ICTTA.2006.1684537
  • Filename
    1684537