• DocumentCode
    462135
  • Title

    High intensity ultrasound phased array for surgical applications

  • Author

    Tan, Meng Tong ; Chu, Chang Min ; Chauhan, Sunita

  • Author_Institution
    Nanyang Technol. Univ., Singapore
  • fYear
    2006
  • fDate
    11-14 Dec. 2006
  • Firstpage
    564
  • Lastpage
    568
  • Abstract
    In this paper, we present a method for simulating the field intensity of ultrasound beam using a two dimensional phased array transducer for High Intensity Focused Ultrasound (HIFU) based surgery. The parameters used and the field modeling is based on Huygen´s principle. We further simulated the field intensity of ultrasound beam in two media, taking into consideration the effect of different attenuation coefficients and beam refraction at the interface. Simulation results show that a focused HIFU beam can be obtained by means of manipulating the phase of the signal in each of the elements in a generic ntimesn-element square array transducer.
  • Keywords
    biomedical transducers; biomedical ultrasonics; surgery; ultrasonic transducers; HIFU based surgery; Huygen´s principle; high intensity focused ultrasound surgery; high intensity ultrasound phased array; ntimesn-element square array transducer; two dimensional phased array transducer; ultrasound beam field intensity simulation; ultrasound beam refraction;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Biomedical and Pharmaceutical Engineering, 2006. ICBPE 2006. International Conference on
  • Conference_Location
    Singapore
  • Print_ISBN
    978-981-05-79
  • Electronic_ISBN
    81-904262-1-4
  • Type

    conf

  • Filename
    4155967