DocumentCode :
462518
Title :
A Gas Proportional Scintillation Counter with KI and KBr-covered Microstrip Plate as Photosensor: Experimental Results
Author :
Carmo, S. J c do ; Carita, L. P M M ; Borges, F.I.G.M. ; Conde, C.A.N.
Author_Institution :
Departamento de Fisica, Univ. de Coimbra
Volume :
3
fYear :
2006
fDate :
Oct. 29 2006-Nov. 1 2006
Firstpage :
1354
Lastpage :
1358
Abstract :
The performance of a xenon-filled gas proportional scintillation counter instrumented with a microstrip plate covered with KI and KBr films as photocathodes to detect the VUV Xe scintillation light is reported. The results for its energy resolution and detector gain for 5.9 keV X-rays are presented and compared with previous results obtained with a CsI film deposited on the microstrip plate under the same conditions. Its performance was studied as a function of the halide film thickness and the scintillation region length. The best energy resolutions obtained reached values of 11.4% and 14.9% for KI and KBr, respectively.
Keywords :
X-ray detection; gas scintillation detectors; photocathodes; position sensitive particle detectors; proportional counters; 5.9 keV; VUV xenon scintillation light; X-rays detection; gas proportional scintillation counter; halide film thickness; microstrip plate photosensor; photocathodes; postassium bromide film; potassium iodide film; scintillation region length; xenon gas; Cathodes; Energy resolution; Instruments; Microstrip; Optical films; Production; Scintillation counters; Solid scintillation detectors; X-ray detection; X-ray detectors;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium Conference Record, 2006. IEEE
Conference_Location :
San Diego, CA
ISSN :
1095-7863
Print_ISBN :
1-4244-0560-2
Electronic_ISBN :
1095-7863
Type :
conf
DOI :
10.1109/NSSMIC.2006.354154
Filename :
4179267
Link To Document :
بازگشت