• DocumentCode
    462682
  • Title

    Image reconstruction from truncated data in SPECT with uniform attenuation

  • Author

    Noo, Frederic ; Defrise, Michel ; Pack, Jed D. ; Clackdoyle, Rolf

  • Author_Institution
    Dept. of Radiol., Utah Univ., Salt Lake City, UT
  • Volume
    4
  • fYear
    2006
  • fDate
    Oct. 29 2006-Nov. 1 2006
  • Firstpage
    2288
  • Lastpage
    2292
  • Abstract
    We prove that accurate SPECT reconstruction of a region-of-interest (ROI) in the activity map is possible from truncated projections collected over 180 degrees, assuming that the attenuation map is constant in a convex region Omega that includes all activity sources. The amount of allowable truncation depends on the size and location of the ROI relative to the direction of the first and last projections and relative to the boundary of Omega. A sketch of the proof is given along with reconstructions from computer-simulated data in support of the theory. Our result may be seen as an extension to SPECT of recent results obtained in classical tomography.
  • Keywords
    image reconstruction; medical image processing; single photon emission computed tomography; ROI; SPECT; attenuation map; classical tomography; image reconstruction; region-of-interest; truncated projections; uniform attenuation; Attenuation; Cities and towns; Collimators; Computed tomography; Image reconstruction; Nuclear and plasma sciences; Nuclear medicine; Radiology; Scattering; Strips; ROI; SPECT; region-of-interest; truncated projections; truncation;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2006. IEEE
  • Conference_Location
    San Diego, CA
  • ISSN
    1095-7863
  • Print_ISBN
    1-4244-0560-2
  • Electronic_ISBN
    1095-7863
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2006.354370
  • Filename
    4179484