    The jPET-D4: Performance evaluation of four-layer DOI-PET scanner using the NEMA NU2-2001 standard

  • Author

    Yoshida, Eiji ; Kobayashi, Ayako ; Yamaya, Taiga ; Watanabe, Mitsuo ; Nishikido, Fumihiko ; Kitamura, Keishi ; Hasegawa, Tomoyuki ; Fukushi, Masahiro ; Murayama, Hideo

  • Author_Institution
    National Inst. of Radiol. Sci., Chiba
  • Volume
    4
  • fYear
    2006
  • fDate
    Oct. 29 2006-Nov. 1 2006
  • Firstpage
    2532
  • Lastpage
    2536
  • Abstract
    We have developed a high-performance brain PET scanner, jPET-D4. This scanner is designed to achieve not only high spatial resolution but also high sensitivity using four-layered depth-of-interaction (DOI) detectors. The scanner has five block detector rings with the ring diameter of 390 mm and each block detector ring consists of 24 DOI detectors. In this paper, we describe the system and performance evaluation of the jPET-D4. In particular, we evaluate scatter fraction, sensitivity and count-rate performance of the jPET-D4 by using the NEMA NU2-2001 standard. The average scatter fraction for this system is 42 % with an energy window of 400-600 keV. The sensitivity at the center of the field-of-view is 19.5 kcps/MBq with energy window of 400-600 keV. Peak NECRs (noise equivalent count rate; NEC1R and NEC2R) with 10 ns coincidence time window are 82 kcps at 8.7 kBq/ml and 56 kcps at 7.6 kBq/ml, respectively. These evaluation measurements indicate the jPET-D4 has excellent performance.
  • Keywords
    brain; image scanners; performance evaluation; positron emission tomography; 10 ns; 390 mm; 400 to 600 keV; NEMA NU2-2001 standard; block detector rings; brain PET scanner; count-rate performance; four-layer DOI-PET scanner; four-layered depth-of-interaction detectors; jPET-D4; noise equivalent count rate; peak NECR; performance evaluation; scatter fraction; Animals; Brain; Crystals; Light scattering; Nuclear and plasma sciences; Positron emission tomography; Solid scintillation detectors; Spatial resolution; Standards development; Whole-body PET;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2006. IEEE
  • Conference_Location
    San Diego, CA
  • ISSN
    1095-7863
  • Print_ISBN
    1-4244-0560-2
  • Electronic_ISBN
    1095-7863
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2006.354425
  • Filename
    4179539