• DocumentCode
    462777
  • Title

    Sampling Requirements for Circular Cone Beam Tomography

  • Author

    Brokish, Jeffrey ; Bresler, Yoram

  • Author_Institution
    InstaRecon, Inc., Urbana, IL
  • Volume
    5
  • fYear
    2006
  • fDate
    Oct. 29 2006-Nov. 1 2006
  • Firstpage
    2882
  • Lastpage
    2884
  • Abstract
    The sampling requirements of tomographic projection data are determined by the set of frequencies occupied by the Fourier transform of the projection data. This region of essential support has been analyzed for various 2-D geometries. The general case of 3-D cone beam projections, using a 2-D detector array, is mainly unexplored. In this paper, we consider the 3-D circular cone-beam (CCB) scan. An analysis of the essential support of CCB projection data provides the sampling requirements for detector spacing, row spacing, and projection count (views).
  • Keywords
    computerised tomography; 3D circular cone-beam scan; 3D cone beam projections; Fourier transform; circular cone beam tomography; Bandwidth; Detectors; Fourier transforms; Frequency; Geometry; Nuclear and plasma sciences; Sampling methods; Sensor arrays; Tomography; USA Councils;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2006. IEEE
  • Conference_Location
    San Diego, CA
  • ISSN
    1095-7863
  • Print_ISBN
    1-4244-0560-2
  • Electronic_ISBN
    1095-7863
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2006.356478
  • Filename
    4179635