DocumentCode
462797
Title
Analytical derivation and experimental verification of a sensitivity formula for slit-slat SPECT collimation
Author
Accorsi, Roberto ; Metzler, Scott D. ; Novak, John R. ; Ayan, Ahmet S. ; Jaszczak, Ronald J.
Author_Institution
Dept. of Radiol., Children´´s Hosp. of Philadelphia, PA
Volume
5
fYear
2006
fDate
Oct. 29 2006-Nov. 1 2006
Firstpage
2976
Lastpage
2980
Abstract
The formula g = wd2/[4pia(d+t)h] sin2thetas is derived for the sensitivity of a slit-slat collimator. Both Monte Carlo and experimental data confirm the expected dependence of g on the parameters, especially the distance from the plane of the slit h and the off-axis angle thetas. To refine agreement, a formula for the sensitivity-equivalent slit width is derived to account for slit penetration. At least at low energy its predictions agree well with Monte Carlo data and its use improves agreement with experimental data.
Keywords
Monte Carlo methods; medical computing; medical image processing; single photon emission computed tomography; Monte Carlo method; sensitivity equivalent slit width; sensitivity formula; slit penetration; slit-slat SPECT collimation; slit-slat collimator; Apertures; Biomedical engineering; Biomedical imaging; Collimators; Detectors; Monte Carlo methods; Nuclear and plasma sciences; Nuclear imaging; Radiology; Single photon emission computed tomography; Collimation; nuclear imaging; pinhole; single photon emission computed tomography (SPECT); slat; slit;
fLanguage
English
Publisher
ieee
Conference_Titel
Nuclear Science Symposium Conference Record, 2006. IEEE
Conference_Location
San Diego, CA
ISSN
1095-7863
Print_ISBN
1-4244-0560-2
Electronic_ISBN
1095-7863
Type
conf
DOI
10.1109/NSSMIC.2006.356500
Filename
4179657
Link To Document