DocumentCode :
462798
Title :
A Compact High Performance Readout Electronics Solution for H9500 Hamamatsu 256 Multianode Photomultiplier Tube for Application in Gamma Cameras
Author :
Popov, Vladimir ; Majewski, Stan
Author_Institution :
Thomas Jefferson Nat. Accelerator Facility, Newport News, VA
Volume :
5
fYear :
2006
fDate :
Oct. 29 2006-Nov. 1 2006
Firstpage :
2981
Lastpage :
2985
Abstract :
We have developed a new readout electronics module for 2-D data readout of H9500 256 (16 times16) Hamamatsu Photomultiplier Tube. This is a 3.04 times 3.04 mm2 anode pad pitch size tube with 1.5 mm thick window. This photo tube is most suitable for high resolution gamma cameras, but the large number of anode outputs makes the readout electronics development complicated. The readout electronics solution to be presented is based on the resistors matrix network (V. Popov, US patent No: 6,747,263 B1) which provides decoupling of all 256 anode outputs into a set of 16 X + 16 Y analog outputs, plus it is integrated with anode gain non-uniformity correction circuit [Popov, et al., 2006], We chose this technique, because it allowed getting a maximum position accuracy performance and significantly reduced the number of analog data acquisition connections. The designed analog circuit uses only standard SMD components, no custom ICs. A small amount of analog outputs allowed to place all electronics components on a small size PCBs and to mount them directly on anode output connectors attached to H9500. This design preserves side-by-side uses of several H9500 PSPMT modules in arrays. We present the results of H9500 readout electronics evaluation in the case of uses of different scintillator pixillated crystal arrays and continuous scintillator crystal.
Keywords :
gamma-ray apparatus; photomultipliers; radioisotope imaging; readout electronics; H9500 Hamamatsu 256 multianode photomultiplier tube; anode gain nonuniformity correction circuit; compact high performance readout electronics solution; continuous scintillator crystal; gamma camera application; readout electronics evaluation; readout electronics module; resistors matrix network; scintillator pixillated crystal array; Anodes; Cameras; Circuits; Data acquisition; Electron tubes; Performance gain; Photomultipliers; Readout electronics; Resistors; Transmission line matrix methods;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium Conference Record, 2006. IEEE
Conference_Location :
San Diego, CA
ISSN :
1095-7863
Print_ISBN :
1-4244-0560-2
Electronic_ISBN :
1095-7863
Type :
conf
DOI :
10.1109/NSSMIC.2006.356501
Filename :
4179658
Link To Document :
بازگشت