Abstract :
A novel charge-metering stimulus amplifier is proposed for high performance neural prosthesis stimulation. The new charge-based approach with feedback has low charge error 0.5%, low charge imbalance 0.2%, and high power efficiency. Charge-metering stimulus amplifier can be implemented using advanced 0.18mum CMOS technology and is compatible with mixed-signal system-on-a-chip (SoC) implantable devices. It also has the capability to monitor the electrode-tissue contact quality and estimate the interface impedance in real-time for robust stimulation.
Keywords :
CMOS integrated circuits; amplifiers; prosthetics; system-on-chip; 0.18 micron; CMOS technology; biomimetic implantable prosthesis; charge-metering stimulus amplifier; neural prosthesis stimulation; system-on-chip; Biomedical electrodes; Biomimetics; CMOS technology; Capacitors; Circuits; Electrical stimulation; Impedance; Prosthetics; Retina; Voltage;