DocumentCode :
46942
Title :
Dual concentric-sectored HIFU transducer with phase-shifted ultrasound excitation for expanded necrotic region: a simulation study
Author :
Jong Seob Jeong
Author_Institution :
Dept. of Med. Biotechnol., Dongguk Univ. - Seoul, Seoul, South Korea
Volume :
60
Issue :
5
fYear :
2013
fDate :
May-13
Firstpage :
924
Lastpage :
931
Abstract :
In high-intensity focused ultrasound (HIFU) surgery, it is desirable to produce a large necrotic area per sonication for reduced treatment time. It has been well known that the conventional split-focus scheme capable of generating multiple foci can increase a necrotic region in the lateral or elevational direction. To treat a deep-seated target, it is necessary to generate an expanded necrotic region in the axial direction. In this paper, a novel sonication scheme capable of producing an expanded coagulated region in the both lateral and axial directions is presented. The proposed method can generate multi-focal spots in the lateral and axial directions by using a dual concentric-sectored (DCS) HIFU transducer based on phase-shifted ultrasound excitation. A sound field simulation was employed for this investigation. Four electrical signals with identical center frequencies and different phases activated the DCS transducer, composed of a disc and an annular element with a confocal point. Four 4-MHz ultrasound signals with different phases were transmitted to the target simultaneously, resulting in generation of dual-focal spots in the lateral and axial directions. The sound field simulation results showed that the ¿6-dB lateral and axial beamwidths of the DCS transducer were maximally 79% and 91% broader than the single-element transducer. Subsequently, bio-heat transfer and thermal dose simulation results were matched to the sound field simulation. Hence, the DCS HIFU transducer combined with phase-shifted excitation may be a promising approach to treat a deep-seated target and to reduce treatment time for HIFU surgery.
Keywords :
biomedical transducers; biomedical ultrasonics; biothermics; blood vessels; dosimetry; surgery; ultrasonic therapy; ultrasonic transducers; axial beamwidths; bioheat transfer; conventional split-focus scheme; deep-seated target; dual concentric-sectored high-intensity focused ultrasound transducer; dual-focal spot generation; electrical signals; elevational direction; expanded coagulated region; expanded necrotic region; frequency 4 MHz; high-intensity focused ultrasound surgery; identical center frequencies; lateral beamwidths; lateral direction; multiple foci generation; novel sonication scheme; phase-shifted ultrasound excitation; single-element transducer; sound field simulation; thermal dose simulation; treatment time; Computer Simulation; High-Intensity Focused Ultrasound Ablation; Humans; Male; Models, Biological; Necrosis; Prostatic Neoplasms; Transducers;
fLanguage :
English
Journal_Title :
Ultrasonics, Ferroelectrics, and Frequency Control, IEEE Transactions on
Publisher :
ieee
ISSN :
0885-3010
Type :
jour
DOI :
10.1109/TUFFC.2013.2649
Filename :
6512829
Link To Document :
بازگشت