• DocumentCode
    469447
  • Title

    The zeus uranium calorimeter - 15 Years of successful operation

  • Author

    Gregor, I.-M.

  • Author_Institution
    Deutsches Elektronen-Synchrotron DESY, Hamburg
  • Volume
    1
  • fYear
    2007
  • fDate
    Oct. 26 2007-Nov. 3 2007
  • Firstpage
    249
  • Lastpage
    253
  • Abstract
    One of the main components of the ZEUS detector at the HERA storage ring is the Uranium Calorimeter (UCAL). It has been running successfully from 1992 when ZEUS started data taking until summer 2007. The UCAL is a uranium-scintillator calorimeter with equal response for electrons and hadrons (e/h = 1.00 plusmn 0.05), a linear energy response and a high energy resolution of sigma(E)/E = 18%/radic(E) for electrons and sigma(E)/E = 35%/radic(E) for hadrons. It covers 99.7% of the solid angle and is able to handle bunch crossing rate of up to 10.4 MHz. This performance demands a very precise calibration and a constant monitoring of the detector. In this paper we present the good performance of this calorimeter over 15 years. Main hardware issues and operating experience are summarised.
  • Keywords
    calibration; nuclear electronics; particle calorimetry; position sensitive particle detectors; readout electronics; scintillation counters; storage rings; uranium; HERA storage ring; ZEUS uranium calorimeter; calibration; calorimeter readout; electrons; hadrons; high energy resolution; uranium-scintillator calorimeter; Calibration; Clocks; Detectors; Electrons; Energy resolution; Helium; Nuclear and plasma sciences; Physics; Poles and towers; Protons; HERA; ZEUS; calibration; uranium scintillator calorimeter;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2007. NSS '07. IEEE
  • Conference_Location
    Honolulu, HI
  • ISSN
    1095-7863
  • Print_ISBN
    978-1-4244-0922-8
  • Electronic_ISBN
    1095-7863
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2007.4436325
  • Filename
    4436325