Calibration strategy for 3-D position sensitive CdZnTe spectrometer arrays
Author :
Feng Zhang ; Zhong He ; Kaye, W.
Author_Institution :
Univ. of Michigan, Ann Arbor
Volume :
1
fYear :
2007
fDate :
Oct. 26 2007-Nov. 3 2007
Firstpage :
459
Lastpage :
463
Abstract :
A new detector system having two layers of three by three 3-D position sensitive CdZnTe gamma-ray spectrometers is being developed. Each detector module consists of a 1.5times1.5times1.0 cm3 pixellated (11times11 pixels) CdZnTe detector and a VAS_UM2.3/TAT4 ASIC front-end board. Using the 3-D position sensing technique, each detector can be electronically divided into 11times11times40 (40 depth slices) small voxels. The detector response within each voxel is reasonably uniform (variations in weighting potential and electron trapping are minimized) and the energy resolution can be very good (~1% FWHM at 662 keV). A calibration matrix can be derived based on the voxel response so that the performance of entire detector can be very close to the performance of small voxels. However, each voxel needs a reasonable amount of events for the calibration, thus the number of calibration events needed for the whole detector is considerably large - ~50 million for each detector. This requirement poses a challenge to the calibration of the system especially for large arrays. Simply reducing the readout dead time for each event may not be the ultimate solution due to other bottlenecks in the system such as data transfer rate and data processing speed. This paper will analyze the bottlenecks in the current dual-array system and discuss possible strategies to speed up the calibration process.
Keywords :
calibration; gamma-ray spectroscopy; nuclear electronics; position sensitive particle detectors; readout electronics; semiconductor counters; 3-D position sensing technique; VAS_UM2.3/TAT4 ASIC front-end board; calibration matrix; data processing speed; data transfer rate; dual-array system; electron trapping; gamma-ray spectrometers; pixellated CdZnTe detector; readout dead time; readout electronics; Application specific integrated circuits; Calibration; Data processing; Electron traps; Energy resolution; Event detection; Gamma ray detection; Gamma ray detectors; Position sensitive particle detectors; Spectroscopy;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Nuclear Science Symposium Conference Record, 2007. NSS '07. IEEE
Conference_Location :
Honolulu, HI
ISSN :
1095-7863
Print_ISBN :
978-1-4244-0922-8
Electronic_ISBN :
1095-7863
Type :
conf
DOI :
10.1109/NSSMIC.2007.4436369
Filename :
4436369
Link To Document :
بازگشت