• DocumentCode
    469793
  • Title

    Calibration of pinhole microSPECT system using line sources

  • Author

    Zhou, Rong ; Ma, Tianyu ; Liu, Yaqiang ; Sun, Xishan ; Kang, Xiaowen ; Wu, Zhaoxia ; Wang, Shi ; Jing, Yongjie ; Cheng, Jianping

  • Author_Institution
    Tsinghua Univ., Beijing
  • Volume
    5
  • fYear
    2007
  • fDate
    Oct. 26 2007-Nov. 3 2007
  • Firstpage
    3364
  • Lastpage
    3367
  • Abstract
    A method using line sources to calibrate a pinhole MicroSPECT system was proposed. Projections\´ coordinates of "vitual" point sources were obtained by solving equations of centerlines, which were detected out of projection images of line sources by Radon transform. Effect of depth of interaction that gamma photons rip into crystals was modeled when calculating the theoretical position of projection. Several systems were simulated by Monte Carlo method and calibrated both by the proposed method and by using ideal point sources for comparison. Calibration accuracy achieved by the proposed method is comparable to that by using ideal point sources.
  • Keywords
    Monte Carlo methods; Radon transforms; calibration; single photon emission computed tomography; Monte Carlo method; Radon transform; calibration; line sources; pinhole MicroSPECT; Calibration; Detectors; Equations; Geometry; Nuclear and plasma sciences; Photonic crystals; Physics; Sun; Telephony; Transforms;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2007. NSS '07. IEEE
  • Conference_Location
    Honolulu, HI
  • ISSN
    1095-7863
  • Print_ISBN
    978-1-4244-0922-8
  • Electronic_ISBN
    1095-7863
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2007.4436853
  • Filename
    4436853