Title :
A non-iterative method for emission tomographic image reconstruction with resolution recovery
Author :
Soon, Victor C. ; Miller, Michael A. ; Hutchins, Gary D.
Author_Institution :
Indiana Univ. Sch. of Med., Indianapolis
fDate :
Oct. 26 2007-Nov. 3 2007
Abstract :
We propose a non-iterative method called Local Regression FBP (LR-FBP) for emission tomographic image reconstruction with resolution recovery. The method processes the sinogram volume data by running a two-dimensional local regression (LR) algorithm over the image projections - this has the effect of improving data signal-to-noise-ratios (SNR) as it fits the image projections thus reducing noise. Resolution recovery is then performed on the fit sinogram and the Filtered Back Projection (FBP) reconstruction using a ramp filter is then applied. The simulation studies presented show that LR-FBP gives lower bias and improved resolution compared to conventional FBP. Imaging studies using the IndyPET-III Small Animal PET scanner and the Siemens/CTI Biograph-16 PET/CT scanner show qualitative improvements in image resolution and contrast, as well as possible quantitative improvements in contrast recovery.
Keywords :
image reconstruction; image resolution; medical image processing; noise; positron emission tomography; IndyPET-III small animal PET scanner; Siemens/CTI Biograph-16 PET/CT scanner; emission tomographic image reconstruction; image projections; image resolution; local regression filtered back projection; noise reduction; noniterative method; ramp filter; resolution recovery; sinogram volume data; two-dimensional local regression algorithm; Animals; Computed tomography; Filters; Image reconstruction; Image resolution; Noise reduction; Positron emission tomography; Signal processing; Signal resolution; Signal to noise ratio;
Conference_Titel :
Nuclear Science Symposium Conference Record, 2007. NSS '07. IEEE
Conference_Location :
Honolulu, HI
Print_ISBN :
978-1-4244-0922-8
Electronic_ISBN :
1095-7863
DOI :
10.1109/NSSMIC.2007.4436877