Abstract :
Time-of-Flight (TOF) PET requires the development of TOF scatter estimation methods. Two different approaches for modeling TOF 3D scatter are presented in this paper. We performed Monte Carlo simulations to compare the two methods. Both methods resulted in good agreement with Monte Carlo simulations. In addition, the second method is independent of the number of timing bins used and therefore, computationally less intensive.
Keywords :
Monte Carlo methods; positron emission tomography; 3D PET; Monte Carlo simulations; TOF 3D scatter modeling; TOF scatter estimation methods; time-of-flight PET; Attenuation; Detectors; Electromagnetic scattering; Gamma rays; Image reconstruction; Integral equations; Nuclear and plasma sciences; Particle scattering; Positron emission tomography; Timing;