• DocumentCode
    469911
  • Title

    Hand-held gamma-ray spectrometer based on high-efficiency Frisch-ring CdZnTe detectors

  • Author

    Cui, Y. ; Bolotnikov, A. ; Camarda, G. ; Hossain, A. ; James, R.B. ; De Geronimo, G. ; Fried, J. ; Connor, P.O. ; Kargar, A. ; Harrison, M.J. ; McGregor, D.S.

  • Author_Institution
    Brookhaven Nat. Lab., Upton
  • Volume
    2
  • fYear
    2007
  • fDate
    Oct. 26 2007-Nov. 3 2007
  • Firstpage
    945
  • Lastpage
    948
  • Abstract
    Frisch-ring CdZnTe detectors have demonstrated both good energy resolution, <1% FWHM at 662 keV, and good efficiency in detecting gamma rays, highlighting the strong potential of CdZnTe materials for such applications. We are designing a hand-held gamma-ray spectrometer based on Frisch- ring detectors at Brookhaven National Laboratory. It employs an 8x8 CdZnTe detector array to achieve a high volume of 19.2 cm3, so that the detection efficiency is greatly improved. By using the front-end application specific integrated circuits (ASICs) developed at BNL, this spectrometer has a small profile and high energy resolution. The spectrometer includes a signal processing circuit, digitization and storage circuits, a high- voltage module, and a universal serial bus (USB) interface. In this paper, we introduce the details of the system structure and report our test results with it.
  • Keywords
    gamma-ray detection; gamma-ray spectrometers; nuclear electronics; semiconductor counters; Brookhaven National Laboratory; Frisch-ring CdZnTe detectors; USB; digitization; energy resolution; front-end application; gamma ray detection; hand-held gamma-ray spectrometer; integrated circuits; signal processing circuit; storage circuits; universal serial bus interface; voltage module; Application specific integrated circuits; Array signal processing; Energy resolution; Gamma ray detection; Gamma ray detectors; Gamma rays; Laboratories; Sensor arrays; Spectroscopy; Voltage; CdZnTe; Frisch-ring detector; gamma-ray spectrometer;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Nuclear Science Symposium Conference Record, 2007. NSS '07. IEEE
  • Conference_Location
    Honolulu, HI
  • ISSN
    1095-7863
  • Print_ISBN
    978-1-4244-0922-8
  • Electronic_ISBN
    1095-7863
  • Type

    conf

  • DOI
    10.1109/NSSMIC.2007.4437174
  • Filename
    4437174