  • Abstract
    High resolution 3D coronary artery MR angiography is time-consuming and can benefit from accelerated data acquisition provided by parallel imaging techniques without sacrificing spatial resolution. Currently, popular maximum likelihood based parallel imaging reconstruction techniques such as the SENSE algorithm offer this advantage at the cost of reduced signal-to-noise ratio (SNR). Maximum a posteriori (MAP) reconstruction techniques that incorporate globally smooth priors have been developed to recover this SNR loss, but they tend to blur sharp edges in the target image. The objective of this study is to demonstrate the feasibility of employing edge-preserving Markov random field priors in a MAP reconstruction framework, which can be solved efficiently using a graph cuts based optimization algorithm. The preliminary human study shows that our reconstruction provides significantly better SNR than the SENSE reconstruction performed by a commercially available scanner for navigator gated steady state free precession 3D coronary magnetic resonance angiography images (n = 4)
  • Keywords
    Bayes methods; Markov processes; angiocardiography; biomedical MRI; graph theory; image reconstruction; image resolution; maximum likelihood estimation; medical image processing; optimisation; Bayesian reconstruction; MAP reconstruction; SENSE reconstruction; SNR loss; data acquisition; edge-preserving Markov random field; graph cuts; high-resolution 3D coronary artery MRI; maximum a posteriori reconstruction techniques; navigator-gated steady state free precession; optimization algorithm; parallel coronary artery magnetic resonance angiography; signal-to-noise ratio; spatial resolution; Angiography; Arteries; Bayesian methods; High-resolution imaging; Image reconstruction; Image resolution; Magnetic resonance; Magnetic resonance imaging; Signal to noise ratio; Spatial resolution;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Engineering in Medicine and Biology Society, 2006. EMBS '06. 28th Annual International Conference of the IEEE
  • Conference_Location
    New York, NY
  • ISSN
    1557-170X
  • Print_ISBN
    1-4244-0032-5
  • Electronic_ISBN
    1557-170X
  • Type

    conf

  • DOI
    10.1109/IEMBS.2006.260300
  • Filename
    4461848