DocumentCode :
471629
Title :
Slotted surface coil with reduced g-factor for SENSE imaging
Author :
Ocegueda, K. ; Rodriguez, A.O.
Author_Institution :
Centre de Imnvestigacion en Imagenologia e Instrum. Medica, Univ. Autonoma Metropolitana, Mexico City
fYear :
2006
fDate :
Aug. 30 2006-Sept. 3 2006
Firstpage :
1904
Lastpage :
1906
Abstract :
A new coil design inspired on the slot-and-hole magnetron tube is proposed for SENSE imaging. To investigate its g-factor behaviour: an SNR formula was derived using the quasi-static approach, and combined with the ultimate g-factor formula to compute the ultimate-factor-g-vs-depth plots. A g-factor expression was derived for the circular coil using the same approach for comparison purposes. SNR-vs-depth profiles of an 4-slot coil showed an important improvement over the circular coil. The 4-slot coil g-factor can be up to 58.32% lower than that of a single circular-shaped coil. This improvement makes the slotted surface coil a good choice for SENSE imaging
Keywords :
biomedical MRI; biomedical equipment; coils; g-factor; SENSE imaging; SNR formula; SNR-vs-depth profiles; circular coil; g-factor behaviour; quasistatic approach; slot-and-hole magnetron tube; slotted surface coil design; Anodes; Cities and towns; Coils; Geometry; Image reconstruction; Magnetic resonance imaging; Noise reduction; Radio frequency; Signal to noise ratio; USA Councils;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society, 2006. EMBS '06. 28th Annual International Conference of the IEEE
Conference_Location :
New York, NY
ISSN :
1557-170X
Print_ISBN :
1-4244-0032-5
Electronic_ISBN :
1557-170X
Type :
conf
DOI :
10.1109/IEMBS.2006.259736
Filename :
4462151
Link To Document :
بازگشت