DocumentCode :
471828
Title :
Optimization of Microelectrode Design for Cortical Recording Based on Thermal Noise Considerations
Author :
Lempka, Scott F. ; Johnson, Matthew D. ; Barnett, David W. ; Moffitt, Michael A. ; Otto, Kevin J. ; Kipke, Daryl R. ; McIntyre, Cameron C.
Author_Institution :
Dept. of Biomed. Eng., Cleveland Clinic Found., OH
fYear :
2006
fDate :
Aug. 30 2006-Sept. 3 2006
Firstpage :
3361
Lastpage :
3364
Abstract :
Intracortical microelectrode recordings of neural activity show great promise as control signals for neuroprosthetic applications. However, faithful, consistent recording of single unit spiking activity with chronically implanted silicon-substrate microelectrode arrays has proven difficult. Many approaches seek to enhance the long-term performance of microelectrode arrays by, for example, increasing electrode biocompatibility, decreasing electrode impedance, or improving electrode interface properties through application of small voltage pulses. The purpose of this study was to use computational models to optimize the design of microelectrodes. We coupled detailed models of the neural source signal, silicon-substrate microelectrodes, and thermal noise to define the electrode contact size that maximized the signal-to-noise ratio (SNR). Model analysis combined a multi-compartment cable model of a layer V cortical pyramidal neuron with a 3D finite element model of the head and microelectrode to define the amplitude and time course of the recorded signal. A spatially-lumped impedance model was parameterized with in vitro and in vivo spectroscopy data and used to define thermal noise as a function of electrode contact size. Our results suggest that intracortical microelectrodes with a contact size of ~380 mum2 will provide an increased SNR in vivo and improve the long-term recording capabilities of silicon-substrate microelectrode arrays
Keywords :
biomedical electrodes; biomedical measurement; electric impedance measurement; finite element analysis; microelectrodes; neurophysiology; prosthetics; thermal noise; 3D finite element model; SNR; Si; chronically implanted silicon-substrate microelectrode arrays; electrode contact size; intracortical microelectrode recording; layer V cortical pyramidal neuron; microelectrode design optimization; model analysis; multicompartment cable model; neural activity; neural source signal; neuroprosthetic application; signal-to-noise ratio; spatially-lumped impedance model; thermal noise; Computational modeling; Design optimization; Electrodes; Impedance; In vivo; Microelectrodes; Neural prosthesis; Signal analysis; Signal to noise ratio; Voltage;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Engineering in Medicine and Biology Society, 2006. EMBS '06. 28th Annual International Conference of the IEEE
Conference_Location :
New York, NY
ISSN :
1557-170X
Print_ISBN :
1-4244-0032-5
Electronic_ISBN :
1557-170X
Type :
conf
DOI :
10.1109/IEMBS.2006.259432
Filename :
4462518
Link To Document :
بازگشت