• DocumentCode
    471829
  • Title

    Analysis of Capacitive Coupling within Microelectrode Array

  • Author

    Hu, Z. ; Troyk, P.R. ; Detlefsen, D.E.

  • Author_Institution
    Biomed. Eng. Dept., Illinois Univ., Chicago, IL
  • fYear
    2006
  • fDate
    Aug. 30 2006-Sept. 3 2006
  • Firstpage
    3365
  • Lastpage
    3368
  • Abstract
    Capacitive coupling within high-density microelectrode arrays can degrade neural recording signal or disperse neural stimulation current. Material deterioration in a chronically implanted neural stimulation/recording system can cause such an undesired effect. We present a simple method with an iterative algorithm to quantify the cross-coupling capacitance, in-situ
  • Keywords
    bioelectric phenomena; biomedical electrodes; iterative methods; microelectrodes; neurophysiology; prosthetics; capacitive coupling; chronically implanted neural stimulation; cross-coupling capacitance; iterative algorithm; microelectrode array; neural recording signal; neural stimulation current; Biomedical electrodes; Biomedical engineering; Biomedical measurements; Capacitance; Capacitors; Coupling circuits; Current measurement; Microelectrodes; USA Councils; Voltage;
  • fLanguage
    English