Abstract
This paper deals with the influence of the echographic probe pressure on the measurement of arterial diameter. It is common sense that arterial diameter measurements are affected by the pressure exerted on the B-mode ultrasound probe but we found no report in the current literature analyzing and quantifying this effect. We demonstrated this influence by the analysis of arterial images obtained with a conventional B-mode ultrasound system while monitoring the pressure exerted on the probe with a force sensor. Our preliminary results demonstrated that continuous measurement was feasible and could be used to assess the relationship between the blood pressure and the arterial diameter. This may prove useful for the quantitative evaluation and the follow-up of patients with cardio-vascular diseases
Keywords
biomedical measurement; blood vessels; cardiovascular system; diameter measurement; echocardiography; force sensors; haemodynamics; probes; B-mode ultrasound measurement; B-mode ultrasound probe; arterial diameter measurement; blood pressure; cardio-vascular diseases; echographic probe pressure; force sensor; pressure monitoring; quantitative evaluation; Biomedical monitoring; Blood pressure; Current measurement; Force sensors; Image analysis; Patient monitoring; Pressure measurement; Probes; Ultrasonic imaging; Ultrasonic variables measurement;
fLanguage
English
Publisher
ieee
Conference_Titel
Engineering in Medicine and Biology Society, 2006. EMBS '06. 28th Annual International Conference of the IEEE
Conference_Location
New York, NY
ISSN
1557-170X
Print_ISBN
1-4244-0032-5
Electronic_ISBN
1557-170X
Type
conf
DOI
10.1109/IEMBS.2006.260219
Filename
4462635
Link To Document