DocumentCode
473748
Title
Reliable automated QT interval measurement for clinical evaluation
Author
Chen, X. ; Xie, W. ; Dong, L. ; Tin, ZK ; Sankaranarayanan, M.
Author_Institution
Inst. for Infocomm Res., Nanyang
fYear
2006
fDate
17-20 Sept. 2006
Firstpage
373
Lastpage
376
Abstract
The aim of this study was to develop a reliable automated method of QT interval measurement for ECG signals, and to participate the PhysioNet/Computers in Cardiology Challenge 2006. A scheme based on Kalman filter was proposed to conduct beat segmentation and searching of the first representative beat from ECG waveform, and a wavelet based algorithm was used to detect the onset of QRS complex and the offset of T wave which yield the measurement ofQT interval. Our preliminary QT measurement results have been submitted to CinC for performance evaluation and further improvement to the algorithms and system is carried out.
Keywords
Kalman filters; electrocardiography; medical computing; performance evaluation; ECG signals; ECG waveform; Kalman filter; QRS complex; T wave; automated QT interval measurement; beat segmentation; clinical evaluation; first representative beat; performance evaluation; wavelet based algorithm; Automatic testing; Cardiology; Costs; Electrocardiography; Gaussian noise; Heart rate interval; Noise measurement; Pollution measurement; Time measurement; Tin;
fLanguage
English
Publisher
ieee
Conference_Titel
Computers in Cardiology, 2006
Conference_Location
Valencia
Print_ISBN
978-1-4244-2532-7
Type
conf
Filename
4511866