• DocumentCode
    473780
  • Title

    Effect of electrocardiogram signal quality on T-wave alternans measurements: A simulation study

  • Author

    Janusek, D. ; Kania, M. ; Maniewski, R.

  • Author_Institution
    Inst. of Biocybernetics & Biomed. Eng., Polish Acad. of Sci., Warsaw
  • fYear
    2006
  • fDate
    17-20 Sept. 2006
  • Firstpage
    505
  • Lastpage
    508
  • Abstract
    In this work a simulation study of T-wave alternans is proposed in order to determine the ability of TWA detectors to quantify the performance characteristics in terms of amplitude estimation accuracy. ECG signals were simulated by repeating a single beat, and by adding four types of noise each at two different levels of power. TWA episodes with different amplitudes and waveforms were added to the signals. In simulated ECG signals sampling frequency as well as amplitude resolution were reduced. We can conclude from this study that all analysed methods of TWA detection are resistant to changes of amplitude resolution but there is strong influence of sampling frequency on the precision of TWA amplitude estimation .
  • Keywords
    biomedical measurement; electrocardiography; medical signal processing; ECG signals; T-wave alternans measurements; TWA detectors; amplitude estimation accuracy; electrocardiogram signal quality; Amplitude estimation; Cardiology; Electrocardiography; Frequency; Sampling methods; Shape; Signal analysis; Signal processing; Signal resolution; Signal sampling;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Computers in Cardiology, 2006
  • Conference_Location
    Valencia
  • Print_ISBN
    978-1-4244-2532-7
  • Type

    conf

  • Filename
    4511899