Estimation of cardiac output from left ventricular pressure by a modified modelflow method
Author
Valsecchi, S. ; Perego, G.B. ; Censi, F. ; Schreuder, J.J.
Author_Institution
Medtronic Italia, Rome, Italy
fYear
2006
fDate
17-20 Sept. 2006
Firstpage
885
Lastpage
888
Abstract
The Modelflow method (MM) requires the recording of the aortic pressure signal, to compute cardiac output (CO) by simulating a non-linear three-element model of aortic input impedance. We propose an extended MM to be applied to the intraventricular pressure signal, to obtain an independent estimation of CO in the setting of conductance measurements. Hemodynamic tests were performed in 21 patients with heart failure during atrio-biventricular pacing. Comparing CO by extended MM simulation with standard MM, the bias was −0.04 (0.36) l/min, with limits of agreement of −0.77 and 0.70 l/min, and a coefficient of variation of 8.4%. In patients without mitral and aortic abnormalities, the continuous CO estimation with the proposed method seems reliable. This method presents an acceptable precision to replace conventional aortic MM estimation in subjects undergoing intraventricular pressure acquisition, such as during conductance catheter evaluation.
Keywords
Blood pressure; Calibration; Catheters; Computational modeling; Electrical resistance measurement; Heart rate; Hemodynamics; Impedance; Performance evaluation; Pressure measurement;