DocumentCode :
473883
Title :
Changes in cardiac indices from implanted defibrillator-stored electrograms due to acquisition and preprocessing conditions
Author :
Requena-Carrión, J. ; Rojo-Álvarez, JL ; Everss, E. ; Alonso-Atienza, F. ; Sánchez-Muñoz, JJ ; Ortiz, M. ; García-Alberola, A.
Author_Institution :
Univ. Carlos III de Madrid, Leganes
fYear :
2006
fDate :
17-20 Sept. 2006
Firstpage :
929
Lastpage :
932
Abstract :
A wide number of cardiac indices have been proposed to describe electrocardiograms (ECG) during Ventricular Fibrillation (VF), and they can be useful when analyzing electrograms (EGM) stored in implantable cardioverter Defibrillator (ICD) during spontaneous VF. However, the dependence of their measurement on acquisition and preprocessing conditions has not been so far statistically quantified. We propose a systematic procedure based on nonparametric bootstrap resampling methods to obtain standard errors and confidence intervals for a test. This test detects changes in the statistical distribution of cardiac indices from ICD-stored EGM during VF, due to discrepancies in acquisition and preprocessing conditions. As an application example, significant changes in the distributions of selected spectral indices due to lead configuration were found by comparing measurements obtained from simultaneously recorded unipolar and bipolar EGM during VF. Our nonparametric bootstrap approach can be readily applied to the measurement of cardiac indices, allowing us to study their changes under a diversity of conditions in a systematic way.
Keywords :
biomedical electrodes; electrocardiography; prosthetics; statistical analysis; ECG; ICD stored EGM; acquisition conditions; bipolar EGM; cardiac index changes; cardiac index statistical distribution; confidence intervals; electrocardiograms; electrogram analysis; implantable cardioverter defibrillator; lead configuration; nonparametric bootstrap resampling methods; preprocessing conditions; spontaneous ventricular fibrillation; standard errors; unipolar EGM; Cardiology; Computer errors; Density measurement; Electrocardiography; Electrodes; Fibrillation; Hospitals; Manufacturing; Measurement standards; Testing;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Computers in Cardiology, 2006
Conference_Location :
Valencia
Print_ISBN :
978-1-4244-2532-7
Type :
conf
Filename :
4512005
Link To Document :
بازگشت