DocumentCode :
48644
Title :
Acoustic Vector-Corrected Impedance Meter
Author :
Scott, James ; Pennington, Kyle E.
Author_Institution :
Sch. of Eng., Univ. of Waikato, Hamilton, New Zealand
Volume :
63
Issue :
12
fYear :
2014
fDate :
Dec. 2014
Firstpage :
2726
Lastpage :
2732
Abstract :
We describe the development of a novel instrument intended for the measurement of the acoustical reflection coefficient of materials. The instrument effectively implements a one-port vector-corrected network analyzer in the acoustic, rather than the electromagnetic, domain. Employing the well-documented methods of error correction familiar to microwave engineers, this instrument permits automated measurement of an acoustic impedance presented to a waveguide port. A dual-directional coupler allows a working frequency range of well over an octave. In principle, a set of six couplers would allow measurement from 100 to 50000 Hz.
Keywords :
acoustic devices; acoustic impedance; acoustic variables measurement; acoustic wave reflection; acoustic waveguides; acoustic impedance; acoustic vector-corrected impedance meter; acoustical reflection coefficient; automated measurement; dual-directional coupler; electromagnetism; error correction; frequency 100 Hz to 50000 Hz; microwave engineers; one-port vector-corrected network analyzer; waveguide port; working frequency; Acoustic measurements; Acoustic reflection; Acoustics; Impedance; Impedance measurement; Acoustic impedance; acoustic measurements; acoustic reflection; impedance measurement; impedance tube; network analyzer; vector correction; vector correction.;
fLanguage :
English
Journal_Title :
Instrumentation and Measurement, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9456
Type :
jour
DOI :
10.1109/TIM.2014.2327474
Filename :
6832511
Link To Document :
بازگشت