System packaging & integration for a swallowable capsule using a direct access sensor
Author :
Jesudoss, Pio ; Mathewson, Alan ; Wright, William ; McCaffrey, Colm ; Ogurtsov, Vladimir ; Twomey, Karen ; Stam, Frank
Author_Institution :
Tyndall Nat. Inst., Cork, Ireland
fYear :
2009
fDate :
15-18 June 2009
Firstpage :
1
Lastpage :
4
Abstract :
Technological developments in biomedical microsystems are opening up new opportunities to improve healthcare procedures. Swallowable diagnostic capsules are an example of this. In this paper, a diagnostic capsule technology is described based on direct-access sensing of the Gastro Intestinal (GI) fluids throughout the GI tract. The objective of this paper is two-fold: i) develop a packaging method for a direct access sensor, ii) develop an encapsulation method to protect the system electronics. The integrity of the interconnection after sensor packaging and encapsulation is correlated to its reliability and thus of importance. The zero level packaging of the sensor was achieved by using a so called Flip Chip Over Hole (FCOH) method. This allowed the fluidic sensing media to interface with the sensor, while the rest of the chip including the electrical connections can be insulated effectively. Initial tests using Anisotropic Conductive Adhesive (ACA) interconnect for the FCOH demonstrated good electrical connections and functionality of the sensor chip. Also a preliminary encapsulation trial of the flip chipped sensor on a flexible test substrate has been carried out and showed that silicone encapsulation of the system is a viable option.
Keywords :
biosensors; encapsulation; flip-chip devices; patient diagnosis; GI tract; anisotropic conductive adhesive interconnect; biomedical microsystems; diagnostic capsule technology; direct access sensor; direct-access sensing; electrical connections; encapsulation method; flexible test substrate; flip chip over hole method; flip chipped sensor; fluidic sensing media; gastro intestinal fluids; healthcare procedures; sensor packaging; silicone encapsulation; swallowable diagnostic capsules; system electronics; system integration; system packaging; zero level packaging; Biosensors; Dielectrics and electrical insulation; Electronics packaging; Encapsulation; Flip chip; Gastrointestinal tract; Intestines; Medical services; Protection; Sensor systems; ACA; Flip chip over hole; direct-access sensor; flexible substrate; polysiloxane;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Microelectronics and Packaging Conference, 2009. EMPC 2009. European
Conference_Location :
Rimini
Print_ISBN :
978-1-4244-4722-0
Electronic_ISBN :
978-0-6152-9868-9
Type :
conf
Filename :
5272917
Link To Document :
بازگشت