DocumentCode :
504485
Title :
Spectral stability of 0.9 µm standard radiation thermometer
Author :
Sakuma, Fumihiro
Author_Institution :
NMIJ, Nat. Inst. of Adv. Ind. Sci. & Technol., Tsukuba, Japan
fYear :
2009
fDate :
18-21 Aug. 2009
Firstpage :
2797
Lastpage :
2801
Abstract :
Long term spectral characteristics of 0.9 mum radiation thermometers were studied by using a monochromator and also fixed point blackbodies. Some thermometers showed shifts in their center wavelengths. Normally the thermometers were stable within plusmn 0.2 nm per year. The trends of the two measurements were the same but they had a small difference in the absolute values. The spectral stability of the radiation thermometers can be monitored by the signal ratio between two fixed point blackbodies.
Keywords :
blackbody radiation; monochromators; spectral methods of temperature measurement; thermometers; fixed point blackbodies; long-term spectral characteristics; monochromator; signal ratio; spectral stability; standard radiation thermometer; wavelength 0.9 mum; Aluminum; Calibration; Copper; Instruments; Interference; Optical filters; Temperature; Thermal stability; Uncertainty; Wavelength measurement; fixed-point blackbodies; interference filter; long-term stability; radiation thermometer; spectral responsivity;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
ICCAS-SICE, 2009
Conference_Location :
Fukuoka
Print_ISBN :
978-4-907764-34-0
Electronic_ISBN :
978-4-907764-33-3
Type :
conf
Filename :
5333555
Link To Document :
بازگشت