• DocumentCode
    504672
  • Title

    Uncertainty for the realization of the Co-C eutectic point for calibration of thermocouples

  • Author

    Ogura, Hideki ; Izuchi, Masaya ; Tamba, Jun ; Arai, Masaru

  • Author_Institution
    Nat. Inst. of Adv. Ind. Sci. & Technol. (AIST), Nat. Metrol. Inst. of Japan (NMIJ), Ibaraki, Japan
  • fYear
    2009
  • fDate
    18-21 Aug. 2009
  • Firstpage
    3297
  • Lastpage
    3302
  • Abstract
    The Co-C cells to realize the metal-carbon eutectic points were constructed, and the uncertainty of the Co-C eutectic point was evaluated using Pt/Pd thermocouples. The uncertainty of Pt/Pd thermocouple calibration at the Co-C eutectic point was estimated to be 0.53 K (k = 2), including the uncertainty of the measuring system and the drift of thermocouple. This result shows the appropriateness of the device for calibrating thermocouples.
  • Keywords
    calibration; cobalt compounds; measurement uncertainty; thermocouples; CoC; measuring system uncertainty; metal-carbon eutectic points; platinum-palladium thermocouples; temperature 0.53 K; thermocouple calibration; thermocouple drift; Calibration; Cobalt; Copper; Filling; Furnaces; Interpolation; Palladium; Powders; Temperature measurement; Uncertainty; Co-C eutectic point; thermocouple calibration; uncertainty;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    ICCAS-SICE, 2009
  • Conference_Location
    Fukuoka
  • Print_ISBN
    978-4-907764-34-0
  • Electronic_ISBN
    978-4-907764-33-3
  • Type

    conf

  • Filename
    5334347