• DocumentCode
    51023
  • Title

    Murine Heart Volume: Numerical Comparison and Calibration of Conductance Catheter Models

  • Author

    Popa, Sorin ; Fricke, Kyle ; Dubois, Jonathan ; Kottam, Anil T. G. ; Sobot, Robert

  • Author_Institution
    Dept. of Electr. & Comput. Eng., Western Univ., London, ON, Canada
  • Volume
    61
  • Issue
    9
  • fYear
    2014
  • fDate
    Sept. 2014
  • Firstpage
    2396
  • Lastpage
    2405
  • Abstract
    A full set of finite-element method (FEM) studies of the catheter within a cylindrical cuvette and within an elliptical cuvette are presented along with novel insight on the fundamental electromagnetic properties of the catheter. An in vitro experiment with modified small mouse pressure-volume catheters was conducted and the results are presented as a validation of the FEM models. In addition, sensitivity analysis on the electrode size and position is conducted and the results allow for a novel calibration factor based on catheter geometry to be presented. This calibration factor is used in conjunction with Wei´s conductance volume equations to reduce the average measured error in cuvette volume measurements from 26.5% to 5%.
  • Keywords
    bioelectric potentials; biomedical electrodes; biomedical measurement; calibration; cardiology; catheters; electric admittance measurement; electromagnetism; finite element analysis; geometry; measurement errors; physiological models; sensitivity analysis; FEM model validation; FEM study; Wei conductance volume equations; average measured error reduction; calibration factor; catheter geometry; conductance catheter model calibration; cuvette volume measurements; cylindrical cuvette; electrode position sensitivity analysis; electrode size sensitivity analysis; electromagnetic properties; elliptical cuvette; finite element method; in vitro experiment; modified small mouse pressure-volume catheters; murine heart volume; numerical comparison; Biomedical measurement; Blood; Catheters; Electrodes; Equations; Mathematical model; Volume measurement; Conductance-based catheter; electromagnetic simulation; finite-element method (FEM); murine heart; pressure??volume; sensitivity analysis; sensor calibration;
  • fLanguage
    English
  • Journal_Title
    Biomedical Engineering, IEEE Transactions on
  • Publisher
    ieee
  • ISSN
    0018-9294
  • Type

    jour

  • DOI
    10.1109/TBME.2014.2313621
  • Filename
    6778025