DocumentCode :
51053
Title :
A Navigation Platform for Guidance of Beating Heart Transapical Mitral Valve Repair
Author :
Moore, J.T. ; Chu, M.W.A. ; Kiaii, B. ; Bainbridge, David ; Guiraudon, G. ; Wedlake, C. ; Currie, Marc ; Rajchl, Martin ; Patel, Rajni V. ; Peters, Terry M.
Author_Institution :
Robarts Res. Inst., Western Univ., London, ON, Canada
Volume :
60
Issue :
4
fYear :
2013
fDate :
Apr-13
Firstpage :
1034
Lastpage :
1040
Abstract :
Traditional surgical approaches for repairing diseased mitral valves (MVs) have relied on placing the patient on cardiopulmonary bypass (on pump), stopping the heart and accessing the arrested heart directly. However, because this approach has the potential for adverse neurological, vascular, and immunological sequelae, less invasive beating heart alternatives are desirable. Emerging beating heart techniques have been developed to offer high-risk patients MV repair using ultrasound guidance alone without stopping the heart. This paper describes the first porcine trials of the NeoChord DS1000 (Minnetonka, MN), employed to attach neochordae to a MV leaflet using the traditional ultrasound-guided protocol augmented by dynamic virtual geometric models. The distance errors of the tracked tool tip from the intended midline trajectory (5.2 ± 2.4 mm versus 16.8 ± 10.9 mm, p = 0.003), navigation times (16.7 ± 8.0 s versus 92.0 ± 84.5 s, p = 0.004), and total path lengths (225.2 ± 120.3 mm versus 1128.9 ± 931.1 mm, p = 0.003) were significantly shorter in the augmented ultrasound compared to navigation with ultrasound alone,1 indicating a substantial improvement in the safety and simplicity of the procedure.
Keywords :
diseases; echocardiography; medical disorders; neurophysiology; surgery; NeoChord DS1000; cardiopulmonary bypass; diseased mitral valves; dynamic virtual geometric models; immunological sequelae; invasive beating heart; neochordae; neurological sequelae; surgery; transapical mitral valve repair; ultrasound-guided protocol; vascular sequelae; Heart; Maintenance engineering; Navigation; Surgery; Trajectory; Ultrasonic imaging; Valves; Image-guided surgery; mitral valve (MV) repair; model-enhanced visualization; off-pump beating heart surgery; Animals; Cardiac Surgical Procedures; Coronary Artery Bypass, Off-Pump; Echocardiography; Mitral Valve; Reproducibility of Results; Surgery, Computer-Assisted; Swine;
fLanguage :
English
Journal_Title :
Biomedical Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
0018-9294
Type :
jour
DOI :
10.1109/TBME.2012.2222405
Filename :
6320686
Link To Document :
بازگشت