DocumentCode :
511348
Title :
A hydrostatic pressure approach to assess accuracy of finger blood pressure measurement
Author :
Raamat, Rein ; Jagomagi, Kersti ; Talts, Jaak ; Kivastik, Jana
Author_Institution :
Department of Physiology, University of Tartu, Estonia
fYear :
2009
fDate :
4-7 Nov. 2009
Firstpage :
1
Lastpage :
4
Abstract :
We studied the feasibility of application of consecutive hydrostatic pressure changes to assess accuracy of measurement of finger blood pressure changes in 11 normal volunteers. The results demonstrated that specific physiological responses to hand postural changes were small in healthy subjects and the consecutive hydrostatic pressure changes of controlled magnitude can be applied to assess accuracy of finger blood pressure measurement. As a calibration checking tool the proposed method may be helpful for cardiovascular diagnostics.
Keywords :
Arteries; Biomedical measurements; Biomedical monitoring; Blood pressure; Fingers; Instruments; Physiology; Pressure measurement; Testing; Wrist; Cardiovascular; accuracy; finger blood pressure; hydrostatic;
fLanguage :
English
Publisher :
ieee
Conference_Titel :
Information Technology and Applications in Biomedicine, 2009. ITAB 2009. 9th International Conference on
Conference_Location :
Larnaka, Cyprus
Print_ISBN :
978-1-4244-5379-5
Electronic_ISBN :
978-1-4244-5379-5
Type :
conf
DOI :
10.1109/ITAB.2009.5394309
Filename :
5394309
Link To Document :
بازگشت