• DocumentCode
    512296
  • Title

    A microfluidic system for semicontinuous bead-based biosensing platforms

  • Author

    Kim, Hye Won ; Kim, Young Man ; Yoo, Sung Keun ; Yoon, Sang Youl ; Lee, Jong Hyun ; Yang, Sung

  • Author_Institution
    Department of Mechatronics, 261 Cheomdan-gwagiro (Oryong-dong), Buk-gu, Gwangju, Korea 500-712
  • Volume
    2009-Supplement
  • fYear
    2009
  • fDate
    2-6 Nov. 2009
  • Firstpage
    1
  • Lastpage
    6
  • Abstract
    Conventional bead-based micro immunoassays do not guarantee high reliability and repeatability due to the trapping mechanisms of the systems which cause inconsistency in trapping of cytometric beads. This paper presents a new beadbased micro immunoassay system for semicontinuous assays with high reliability and accuracy in cytometric bead trapping by using the hole-patterned microstructures. Trapping test showed the assembly of single layer of bead array in ordered fashion. In addition that, the µ-immunoassay exhibits that the fluorescence intensity linearly increases in proportion to Biotin-4-Fluorescin (B4F) concentration and a 10 pg/mL appears to be the lowest detectable concentration for B4F after incubation time of 5 minutes.
  • Keywords
    Biosensors; Microfluidics; Bead-based microimmunoassay; cytometric bead trapping; microfluidic device;
  • fLanguage
    English
  • Publisher
    ieee
  • Conference_Titel
    Communications and Photonics Conference and Exhibition (ACP), 2009 Asia
  • Conference_Location
    Shanghai, China
  • Print_ISBN
    978-1-55752-877-3
  • Electronic_ISBN
    978-1-55752-877-3
  • Type

    conf

  • Filename
    5405578