DocumentCode :
51669
Title :
Application of Implantable Wireless Biomicrosystem for Monitoring Nerve Impedance of Rat After Sciatic Nerve Injury
Author :
Yu-Ting Li ; Chih-Wei Peng ; Lung-Tai Chen ; Wen-Shan Lin ; Chun-Hsun Chu ; Chen, J.J.
Author_Institution :
Dept. of Biomed. Eng., Nat. Cheng Kung Univ., Tainan, Taiwan
Volume :
21
Issue :
1
fYear :
2013
fDate :
Jan. 2013
Firstpage :
121
Lastpage :
128
Abstract :
Electrical stimulation is usually applied percutaneously for facilitating peripheral nerve regeneration. However, few studies have conducted long-term monitoring of the condition of nerve regeneration. This study implements an implantable biomicrosystem for inducing pulse current for aiding nerve repair and monitoring the time-course changes of nerve impedance for assessing nerve regeneration in sciatic nerve injury rat model. For long-term implantation, a transcutaneous magnetic coupling technique is adopted for power and data transmission. For in vivo study, the implanted module was placed in the rat´s abdomen and the cuff electrode was wrapped around an 8-mm sciatic nerve gap of the rat for nerve impedance measurement for 42 days. One group of animals received monophasic constant current via the cuff electrode and a second group had no stimulation between days 8-21. The nerve impedance increased to above 150% of the initial value in the nerve regeneration groups with and without stimulation whereas the group with no nerve regeneration increased to only 113% at day 42. The impedance increase in nerve regeneration groups can be observed before evident functional recovery. Also, the nerve regeneration group that received electrical stimulation had relatively higher myelinated fiber density than that of no stimulation group, 20686 versus 11417 fiber/mm 2. The developed implantable biomicrosystem is proven to be a useful experimental tool for long-term stimulation in aiding nerve fiber growth as well as impedance assessment for understanding the time-course changes of nerve regeneration.
Keywords :
bioelectric phenomena; biomedical electrodes; biomedical measurement; electric impedance measurement; fibres; injuries; neurophysiology; patient monitoring; prosthetics; 20686-11417 fiber; abdomen; cuff electrode; data transmission; electrical stimulation; high myelinated fiber density; implantable wireless biomicrosystem; implanted module; long-term monitoring; monophasic constant current; nerve fiber growth; nerve impedance measurement; nerve regeneration groups; peripheral nerve regeneration; power transmission; pulse current; sciatic nerve injury rat model; transcutaneous magnetic coupling technique; Educational institutions; Electrical stimulation; Electrodes; Impedance; Impedance measurement; Monitoring; Nerve fibers; Electrical stimulation; impedance measurement; implantable biomicrosystem; nerve regeneration; Animals; Electrodes, Implanted; Equipment Design; Equipment Failure Analysis; Male; Miniaturization; Monitoring, Ambulatory; Plethysmography, Impedance; Prostheses and Implants; Rats; Rats, Wistar; Sciatic Neuropathy; Telemetry;
fLanguage :
English
Journal_Title :
Neural Systems and Rehabilitation Engineering, IEEE Transactions on
Publisher :
ieee
ISSN :
1534-4320
Type :
jour
DOI :
10.1109/TNSRE.2012.2219883
Filename :
6323039
Link To Document :